3
face velocities, both methods at each
First
Tri-Color flagged the horizontal openings before tracer gas was run there
0.00 ppm
tracer gas at 70 fpm, at the spot where Tri-Color showed escape
The ask
The number, and the picture
The lab's lead engineer needed as-installed ASHRAE 110 tracer gas results and wanted to see dynamic containment testing next to it. We ran Tri-Color beside tracer gas on an 8 ft bench hood with a combination sash on VAV controls.
How it worked
Same hood, same day
Tri-Color's laser curtain at the sash plane, the mannequin and probe on the same hood. Both ran at 100 fpm as installed. Then the controls contractor turned the hood down to 50 fpm and back up to 70 fpm, each setpoint confirmed at the opening, and we ran both again.
What Tri-Color showed
Where it leaks, and when
At 100 fpm the red laser dipped low at the horizontal panel openings; tracer gas then failed there on a 0.27 ppm peak against the spec's 0.10 ppm. At 50 fpm both failed. At 70 fpm Tri-Color showed escape at the left side while tracer gas at that spot read 0.00 ppm.
Side by side
Same spot, two different answers
| Condition | Tri-Color | SF₆ tracer gas | Together |
|---|---|---|---|
| 100 fpm, vertical sash | No loss of containment | 0.00 ppm avg, 0.00 peak at left, center, right. Pass | Agree |
| 100 fpm, horizontal panels | Red laser dipping low at the openings, a precursor to loss | 0.27 ppm peak at center. Fail | Tri-Color flagged it first |
| 50 fpm, left, vertical sash | Clear, sustained loss at the left side | 0.12 ppm avg, 0.75 peak. Fail | Both fail |
| 70 fpm, left, vertical sash | Persistent loss at the left side | 0.00 ppm avg, 0.00 peak. Pass | Tri-Color sees what the probe doesn’t |


The 50 and 70 fpm runs were below the hood’s installed setpoint, outside its as-installed rating.

Two methods, two questions
Tracer gas tells you how much reached one point: a still mannequin’s breathing zone. Tri-Color shows whether anything crosses the sash plane, anywhere along the opening, in real time, with people working at the hood. When the loss was big enough to reach the probe, at 50 fpm, both agreed, so Tri-Color isn’t raising false alarms. It sees the same event earlier and across the whole opening.
Tracer gas gives you a number to hold against a spec. Tri-Color shows whether the hood holds, and where to look when it doesn’t.
That isn’t a fault in tracer gas. It’s a limit of one probe at one point.
From the test